54
Chapter 5 Operation
VISERA XENON LIGHT SOURCE CLV-S40
5.3 Adjusting the brightness (automatic/manual)
If the endoscopic image dims during use, this may be a sign
that blood or mucus is adhered to the light guide on the
endoscope’s distal end. Remove the blood or mucus in order
to obtain optimum illumination and to ensure the safety of
examination. If you continue to use the endoscope in such a
condition, the distal end temperature may rise and cause
injury to the patient and/or operator.
Automatic brightness control 1.
When this instrument is combined with any of the following products, be sure to
use the automatic brightness control.
OTV-S5
OTV-S6
OTV-SX
OTV-SX2
Video system that uses a BNC cable
Combination of the OTV-S7V and a VISERA videoscope
(ENF TYPE V, CYF TYPE V, CYF TYPE VA or HYF TYPE
V)
Before disconnecting the video system from the endoscope,
be sure to press the standby switch to disable the light, or
minimize the brightness using manual brightness control. If
the video system is disconnected without performing one of
these operations, the light may cause operator eye injury or
burns.
When switching the normal intensity mode to high intensity
mode, be sure to set the brightness level to below 0.
Otherwise, the brightness will exceed the necessary level. It
may result operator and/or patient injury.