Chapter 3 Installation and Connection
19
VISERA XENON LIGHT SOURCE CLV-S40
Chapter 3 Installation and Connection
In the event of instrument failure, or malfunction, always have
another light source available for immediate use.
Turn the light source OFF, and confirm that the power
switches for all other equipment are “OFF” before connecting
them. Otherwise, equipment damage can result.
Use only light guide cables that are compatible with this
instrument. Equipment damage or malfunction can result if
non-compatible cables are used.
Do not place the light source in a location exposed to direct
sunlight, X-rays, radioactivity or strong electromagnetic
radiation (near microwave medical treatment equipment,
short-wave medical treatment equipment, MRI or radio
equipment). The light can be unstable.
Prepare the instrument and other equipment that are used with this instrument
before each use, according to the “System chart” in the Appendix. Refer to the
instruction manuals for each equipment. Install and connect all equipment
described as follows:
3.1 Installation
Do not place any equipment other than the Olympus video
system on the top of the light source. Otherwise, equipment
damage can result.
The light source’s air vents should be clear of ancillary
equipment. Blockage can cause overheating and equipment
damage.
Place the light source on a stable level surface, using the foot
holders (MD-512).
If the instrument is to be placed on a cart, make sure that the
cart is of adequate size and strength to support it.