EasyManua.ls Logo

Olympus VISERA CLV-S40 - Video System Compatibility Details; Powering On and Igniting the Lamp

Olympus VISERA CLV-S40
82 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 5 Operation
53
VISERA XENON LIGHT SOURCE CLV-S40
Fiberscope
CHF TYPE CB30L/S,
URF TYPE P3
Videoscope
LTF TYPE V/V2/V3,
A4941A/43A,
A4801A/03A/05A,
A50001A/03A/21A/23A
This instrument includes a memory function for the intensity
mode, which stores the intensity setting being used when the
light source is turned OFF. Therefore, the user may start the
next operation with the same setting used when the light
source was turned OFF.
The default setting of this function is normal intensity.
5.2 Power supply and igniting the lamp
When turning ON the light source, never to allow the distal
end of an endoscope to come in contact with patient,
operating room drapes, and/or other flammable material.
Patient injury and/or fire can result.
Turn the light source ON and connect the endoscope to the output connector on
the front panel (see Figure 5.1).
Figure 5.1

Related product manuals