Chapter 7 Care, Storage and Disposal
65
VISERA XENON LIGHT SOURCE CLV-S40
Chapter 7 Care, Storage and Disposal
7.1 Care
Use only a dry lint-free cloth to clean the power switch. Do
not moisten the lint-free cloth; the switch could fail or electric
shock may result.
After cleaning the light source, dry it thoroughly before use. If
it is used while still wet, there is the risk of an electrical shock.
When cleaning the light source, always wear personal
protective equipment, such as eye wear, a face mask,
moisture-resistant clothing and chemical-resistant gloves that
fit properly and are long enough so that your skin is not
exposed. Blood, mucous and other potentially infectious
material adhering to the light source could pose an infection
control risk.
Do not clean the output connector, other connectors or the
AC inlet. Cleaning them can deform or corrode the contacts
and/or cause equipment damage.
Do not autoclave or gas sterilize the light source. Equipment
damage will occur.
Do not wipe the outside surface of the light source with an
abrasive material. The surface of the unit will be scratched.
If the light source becomes soiled, perform the following cleaning procedure
immediately after use. If cleaning is delayed, organic debris will solidify and it
may be difficult to effectively clean the light source. The equipment should be
cleaned routinely.
1. Turn the light source OFF and disconnect the power cord.
2. Should the equipment become soiled with blood or other potentially
infectious materials, first wipe off all gross debris using a soft, lint-free cloth
moistened with detergent solution.
3. Wipe the outside of the unit using a soft, lint-free cloth moistened with 70%
ethyl or isopropyl alcohol to remove dust, dirt, etc.
4. Dry the light source with a clean, lint-free cloth.