36
Chapter 4 Inspection
VISERA XENON LIGHT SOURCE CLV-S40
When the warning buzzer does not sound
If the emergency lamp indicator blinks but the warning buzzer does not sound,
an emergency lamp malfunction may be suspected. Stop using the light source
and contact Olympus.
4.2 Checking the lamp life
After turning the light source ON, confirm that the indicator in the lamp life meter
on the front panel of the light source is between 0 and 500 hours (see Figure
4.4).
The lamp life meter shows how many hours the lamp has been used.
Figure 4.4
If the indicator reaches over 500h, replace the lamp with a new one as described
in Chapter 6, “Lamp Replacement”.
Replace an expired lamp with a new one as soon as
possible. An expired lamp may explode or go out during use.
If a lamp explosion occurred, contact Olympus.
• 0 – 300h
One scale unit: 50h
• 300 – 500h
One scale unit: 100h
Lamp life meter