Chapter 4 Inspection
35
VISERA XENON LIGHT SOURCE CLV-S40
When a circuit breaker trips, do not reset it immediately. Turn
the light source OFF and wait at least 15 seconds before
resetting the circuit breakers to prevent damaging them.
4. Turn the light source OFF, disconnect the power cord from the wall mains
outlet and depress both the circuit breakers until they click (see Figure 4.3).
Figure 4.3
5. Reconnect the power cord and turn the light source ON. Confirm that power
is being supplied to the light source.
If the emergency lamp indicator goes on
 When the warning buzzer sounds
1. If the emergency lamp indicator blinks and the warning buzzer sounds,
check to see if the emergency lamp switch lever is properly set to the main
lamp position. Unless the switch lever is set to this position, the main lamp
(xenon) will not light. If the emergency lamp switch lever is not properly set
to the main lamp position, move the switch lever all the way to this position.
Then, after turning the light source OFF, turn the light source ON again to
ignite the xenon lamp.
2. When the light source is turned ON, it will take approximately 5 seconds
before the main lamp (xenon) lights up. If it does not come on, turn the light
source OFF again and then ON again to repeat the lamp ignition. If the lamp
still fails to ignite, replace it with a new one.
Circuit breakers