34
Chapter 4 Inspection
VISERA XENON LIGHT SOURCE CLV-S40
When this instrument and the VISERA video system center
OTV-S7V are combined with any of the following products,
the brightness control indicator and MANUAL indicator are
disabled.
OTV-S7V camera head
LTF TYPE V3
A50001A/03A/21A/23A
If power fails to come on
If power fails to come on, inspect the system as follows:
1. Confirm that the power cord is connected securely to a wall mains outlet and
the AC inlet on the light source.
2. Confirm that the lamp access cover is completely closed (see Figure 4.2).
If it is not closed, power cannot be supplied to the light source.
Figure 4.2
3. Confirm the circuit breakers. If the breakers are not popped out, turn the
light source OFF. Do not use the light source and immediately contact
Olympus. If the breakers have tripped (popped out), reset them according to
steps 4. and 5..
Reset the circuit breakers only once. If power is still not
supplied to the light source after resetting the circuit
breakers, immediately turn the light source OFF, disconnect
the power cord from the wall mains outlet and contact
Olympus. This condition can result in electrical shock and/or
fire.
Turn knob to lock mark.
Lock mark
Lamp access cover