Connections and Interfaces
39
5.2.7.2 LL out 1-6 (Low Level Outputs 1-6)
The SELV interface connector "LL out 1 - 6" holds two independent
generator triples. These six high accuracy analog signal sources can serve
to either control an external amplifier or to directly provide small signal
outputs.
In addition, a serial digital interface is available that transmits control and
monitor functions between the CMC 356 and the external amplifiers.
Supported devices are CMA 156, CMA 56
1
, CMS 156, CMS 251
1
and
CMS 252
1
.
The low level outputs are short-circuit-proof and continually monitored for
overload.
Connect the external amplifier to the CMC 356 low level outputs. Use the
connecting cable that was supplied with the amplifier.
For more detailed information please refer to the technical data section
6.3.5, "Low Level Outputs "LL out" for External Amplifiers" on page 54.
5.2.7.3 LL out 7-12 (Low Level Outputs 7-12) - Option "LLO-2"
The SELV interface connector "LL out 7 - 12" is optionally available for the
CMC 356 test set.
The outputs 7-12 extend the low level outputs 1-6 by two more independent
generator triples. Outputs 7-12 are technically identical to outputs 1-6 as
described above.
For more detailed information please refer to the technical data section 6.12,
"Option LLO-2 (Low Level Outputs)" on page 94.
Overload Warning Flagged in the Software
When a low level output is overloaded, a corresponding warning message
appears on the user interface of the OMICRON Test Universe software.
1
These products are not available anymore.