6TECHNICAL DATA
Guaranteed Values:
General:
The values are valid for the period of one year after factory calibration,
within 23 °C ± 5 °C at nominal value and after a warm-up time greater
than 25 min.
Guaranteed values from the generator outputs:
The values are valid in the frequency range from 10 to 100 Hz unless
specified otherwise. Given maximum phase errors are related to the
voltage amplifier outputs.
Accuracy data for analog outputs are valid in the frequency range from
0 to 100 Hz unless specified otherwise.
The given input/output accuracy values relate to the range limit value
(% of range limit value).
6.1 Main Power Supply
Table 6-1:
Power supply data
Main Power Supply
Connection Connector according to IEC 60320-1 C14
Voltage, single phase
nominal voltage
operational range
100-240V
AC
85 ... 264 V
AC
Power fuse T 12.5 AH 250 V (5 x 20 mm)
"Schurter", order number 0001.2515
Nominal current
1
1
Refer to section 6.3.4, "Operational Limits in Conjunction with Mains Supply" on page 53.
at < 170 V: 12 A max.
at > 170 V: 10 A max.
Frequency
nominal frequency
operational range
50/60 Hz
45 ... 65 Hz
Overvoltage category II