Technical Data
81
6.11.6.3 Accuracy of Phase Measurement
Figure 6-23:
Phase error as function of
input voltage
Conditions: integration time 1 s, measurement signal sinusoidal,
measurement range 100 V, f = 50 Hz, sampling rate 28.44 kHz.
Figure 6-24:
Phase error as function of
sampling rate
Conditions: integration time 1 s, measurement signal sinusoidal, f = 50 Hz,
measurement range 100 V, both channels same excitation (20 V, 70 V).
Phase Error as a Function of the input voltage
(Phase CH1-CH2, Range:100 V, f = 50 Hz)
0
0,05
0,1
0,15
0,2
0,25
0,3
110100
Amplitude CH2 (Vrms)
Phase Error / °
CH1:10V
CH1:70V
CH1:50V
Phase. error in °
Phase error as function of the input voltage
Phase CH1-CH2; range: 100 V; f = 50 Hz
Amplitude CH2 in Vrms
Phase Error as function of the sampling rate
(fin = 50Hz, R:100V)
0
,