EasyManua.ls Logo

Parker NitroFlow HP - Calibrate Oxygen Sensor

Parker NitroFlow HP
59 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
K3.1.141e Manual NitroFlow HP_EN.doc - 50 -
9.5 Calibrate oxygen sensor
1. Stop the unit by touching the
button in the main screen (refer to §6.2.1).
2. Let the system depressurize
Warning:
Do not touch live wires!
3. Remove the sensor cap (E) and expose the sensor to ambient air.
4. Enter the maintenance menu
screen 3/5 and wait until the oxygen level has
stabilized (this will take approx. 60 seconds).
5. Go to screen 5/5 and touch O2 20.9% (span) button. The system will ask for
confirmation: select YES.
6. Start the NitroFlow
®
HP by touching the button in the main screen (refer to
§6.2.1).
7. Check the sample flow of the sensor (0.3 l/min.) coming out of the tube (F) that is
connected to the sensor cap (E). As an indication the sample flow is only just felt
when flown against a wet skin.
8. Reconnect the tube (F) and the sensor cap (E) with the sensor (C).

Table of Contents

Related product manuals