11 Upgrades
218
Option B72 (FM30 only)
1 Remove the top cover assembly as described “Removing the Top Cover Assembly” on page 141.
2 Remove the noninvasive blood pressure assembly.
3 Remove the front end assembly -- see “Removing the Front End Assembly” on page 200.
4 Remove the connector block frame from the front end assembly (see “Disassembling the Front
End Assembly” on page 202). Leave the fetal sensor socket connector block in place.
5 Remove the blood pressure connector from the bus master board.
6 Join the new SpO
2
connector to the noninvasive blood pressure connector.