Transducer Safety Transducers
Lumify Ultrasound System 115
Philips Healthcare 4535 618 58571_A/795 * MAY 2016
Transducers
The transducer is the most important factor in image quality. Optimal imaging cannot be
obtained without the correct transducer. The system is optimized for use based on your
transducer selection.
Transducer Safety
WARNING
Use only Philips transducers and Philips-approved biopsy guides, covers, brackets, supplies,
components, and accessories. Other brands may not properly fit Philips transducers.
Improper installation may result in patient injury.
WARNING
Always remove the transducer from the patient before defibrillation.
WARNING
Before defibrillation, if you cannot remove the transducer from the patient, always
disconnect invasive transducers that remain in contact with the patient from the system.
WARNING
To limit potential harm when scanning neonatal, pediatric, and medicated patients,
minimize the time spent imaging at temperatures above 41°C (106°F).