EasyManua.ls Logo

Philips Lumify - Adding a Modality Worklist Server

Philips Lumify
168 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Modality Worklist Using the System
Lumify Ultrasound System 87
Philips Healthcare 4535 618 58571_A/795 * MAY 2016
Adding a Modality Worklist Server
1. Do one of the following:
Touch , and then touch Setup Modality Worklist .
On the Patient Info form, touch Query MWL , and then touch Setup .
NOTE
If you have previously added a Modality Worklist server, Query MWL will query the server,
instead of letting you access Setup.
2. In Setup Worklists, touch Add New.
3. In the Setup MWL Server dialog box, type or select values from the menus for the
following:
Server Nickname
Device AE Title (the AE title for your device)
Peer AE Title (the AE title for the Modality Worklist server)
Hostname or IP (use a DNS or a static IP address)
DNS Suffix
Port (the port number for the Modality Worklist server)
4. Select Query Options.
5. To specify advanced connection settings, select Show Advanced Options:
DNS Suffix: The DNS name without the hostname
Read Timeout (Sec): The network reply timeout
Connection Timeout (Sec): The DICOM ARTIM timeout
6. To test the connection to the server, touch Test.

Table of Contents