System Maintenance Error Messages
154 Lumify Ultrasound System
Philips Healthcare 4535 618 58571_A/795 * MAY 2016
Troubleshooting
Symptom Corrective Action
The system does not power up. Verify that the device is fully charged.
The system spontaneously reverts to the Scan/Create
Patient display.
Verify that the device is fully charged.
The system does not recognize a connected transducer. Disconnect the ultrasound USB cable from the
transducer and connect a standard Type A to Micro B
USB cable. Connect the temporary cable and the
transducer to a Windows PC. Open the Device Manager.
If the transducer is working properly, PiUsb appears in
Other Devices. If you do not see PiUsb, contact your
Philips representative for a replacement transducer or
cable.
The system continuously reinitializes the transducer
when attempting to image.
Verify that the device is fully charged.
Registration fails. Make sure that you have constant wireless or cellular
network connectivity throughout the registration
process, and make sure that the transducer cable is
firmly connected to your device.
Image artifacts appear. Run the transducer test. See “Testing Transducers” on
page 123.
Error Messages
The system displays error messages in response to operating or error conditions detected by
the system.
The error messages must be noted and reported to your Philips representative.