Electromagnetic Compatibility Safety
Lumify Ultrasound System 57
Philips Healthcare 4535 618 58571_A/795 * MAY 2016
Approved Transducers for Electromagnetic Compliance
The imaging transducers used with the system may affect its emissions. The transducers listed
in “Clinical Applications and Transducers” on page 116, when used with the system, have been
tested to comply with the Group 1, Class A emissions, as required by international standard
CISPR 11. Use only those transducers.
WARNING
Using cables, transducers, and accessories other than those specified for use with the
system may result in increased emissions from, or decreased immunity of, the system.
Approved Accessories for Electromagnetic Compliance
Accessories used with the system may affect its emissions. The accessories listed here, when
used with the system, have been tested to comply with the Group 1, Class A emissions as
required by International Standard CISPR 11. Use only the accessories listed here.
When connecting other accessories to the system, such as a printer or computer, it is the user’s
responsibility to ensure the electromagnetic compatibility of the system. Use only CISPR 11 or
CISPR 22, Class A- or B-compliant devices, unless otherwise noted.
WARNING
Using cables, transducers, and accessories other than those specified for use with the
system may result in increased emissions from, or decreased immunity of, the system.