Transducer Care Transducer and Cable Cleaning
132 Lumify Ultrasound System
Philips Healthcare 4535 618 58571_A/795 * MAY 2016
The care method for your transducer determines the appropriate disinfectant for your
transducer. For details about compatible disinfectants, see “Disinfectants Compatibility” on
page 140.
Transducer and Cable Cleaning
These general cleaning instructions are indicated for all non-TEE transducers, cables, and
connectors. It is important that you clean the transducer, cable, and USB connector according
to the following procedures.
Before cleaning a transducer, read the “Safety” section and “Transducer Care and Operator
Safety” on page 125.
After cleaning, you must disinfect or sterilize non-TEE transducers by following the appropriate
procedures:
“Low-level Disinfecting of Transducers” on page 135
“High-level Disinfecting of Transducers” on page 137
WARNING
Always use protective eyewear and gloves when cleaning and disinfecting any equipment.
CAUTION
When cleaning and disinfecting transducers, do not allow any fluid to enter electrical
connections or metal portions of the USB connector. Damage due to fluids in these areas is
not covered by the warranty or your service contract.