High-level Disinfecting of Transducers Transducer Care
Lumify Ultrasound System 139
Philips Healthcare 4535 618 58571_A/795 * MAY 2016
Immersion Point for USB Transducers
1 Immerse this portion only.
2 Maximum allowable immersion depth; you are not required to immerse to this depth if it is unnecessary.
Minimizing the Effects of Residual Disinfectant
If you use an OPA (ortho-phthalaldehyde) based disinfectant, residual solution may remain on
your transducers if you do not carefully follow the manufacturer’s instructions.
To minimize the effects from residual OPA, or any other disinfectant, Philips recommends the
following:
• Follow the disinfectant manufacturer’s instructions very carefully. For example, the
manufacturer of Cidex OPA recommends rinsing transducers by immersing them three
times in fresh water.