Electrical Safety Safety
Lumify Ultrasound System 27
Philips Healthcare 4535 618 58571_A/795 * MAY 2016
CAUTION
For information on electromagnetic emissions and immunity as it applies to the system, see
“Electromagnetic Compatibility ” on page 53. Ensure that the operating environment of
your system meets the conditions specified in the referenced information. Operating the
system in an environment that does not meet those conditions may degrade system
performance.
Defibrillators
Observe the following warnings when a defibrillation is required while using the ultrasound
system.
WARNING
Before defibrillation, always remove all patient-applied parts from the patient.
WARNING
Before defibrillation, always disconnect invasive transducers that remain in contact with the
patient from the system.
WARNING
A disposable transducer cover provides no protective electrical insulation against
defibrillation.