Safety Electromagnetic Compatibility
56 Lumify Ultrasound System
Philips Healthcare 4535 618 58571_A/795 * MAY 2016
Electromagnetic Emissions: Environment Guidance
Emissions Test Compliance Electromagnetic Environment
Guidance
RF emissions, CISPR 11 Group 1 The system uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
RF emissions, CISPR 11 Class A The system is suitable for use in all
establishments, except domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions, IEC 61000-3-2 Class A
Voltage fluctuations/flicker
emissions, IEC 61000-3-3
Complies
Approved Cables for Electromagnetic Compliance
Cables connected to the system may affect its emissions. Use only the cable types and lengths
listed here.
WARNING
Using cables, transducers, and accessories other than those specified for use with the
system may result in increased emissions from, or decreased immunity of, the system.
Approved Cables
Cable Length Philips Part
Number
Transducer Cable Kit (C5-2 , L12-4) 1.25 m (4 ft) 453561806941