Testing Transducers Transducers
Lumify Ultrasound System 123
Philips Healthcare 4535 618 58571_A/795 * MAY 2016
Daily and Long-Term Storage
Follow these guidelines to protect your transducer:
• Avoid storing transducers in areas of temperature extremes or in direct sunlight.
• Store transducers separately from other instruments to avoid inadvertent transducer
damage.
• Before storing transducers, make sure they are thoroughly dry.
Testing Transducers
You can run transducer tests to diagnose image quality and transducer issues.
1. Make sure that your device is connected to a wireless or cellular network.
2. Connect the transducer to your device.
3. Make sure that the transducer lens is clean, dry, and not touching anything.
4. Touch and select Settings .
5. In Transducer Tests, touch Run Tests.
The system runs a series of tests and then sends the logs to Philips Remote Services. If your
device is not connected to a wireless or cellular network, the logs are queued until you have
network connectivity. For more information, contact your Philips representative or visit the
Lumify portal:
www.philips.com/lumify