Performing an Exam Measurement Accuracy Tables
98 Lumify Ultrasound System
Philips Healthcare 4535 618 58571_A/795 * MAY 2016
WARNING
System users are responsible for image quality and diagnosis. Inspect the data that is being
used for the analysis and diagnosis and ensure that the data is sufficient both spatially and
temporally for the measurement approach being used.
Measurement Accuracy Tables
2D Measurement Range and Accuracy
Measurement Accuracy Maximum Range
Axial Distance ≤ ± 2% or 2 mm >30.0 cm
Lateral Distance ≤ ± 2.5% or 3 mm >40.0 cm
Diagonal Distance ≤ ± 2% or 2 mm >32.0 cm
Acquiring Images
You can acquire and save a still image from the current exam. The acquired image is saved in
the patient exam, and a thumbnail of it is available in the Review display.
Acquiring an Image in Live Imaging
Touch Save Image . The system beeps when the image acquisition is complete.
Acquiring an Image in Live Imaging to Measure or Annotate
Touch , and then do any of the following:
• To select a frame to acquire, drag the scroll bar or touch or to move through the
images in the cineloop sequence to display the image that you want to acquire.