Basic Safety Safety
Lumify Ultrasound System 23
Philips Healthcare 4535 618 58571_A/795 * MAY 2016
WARNING
Stop use immediately if the system or the transducer appear to be malfunctioning. Contact
your Philips representative immediately.
WARNING
You are responsible for configuring your device in accordance with your institution's security
policies. Notifications and alerts from third-party applications may interfere with an exam.
WARNING
Thin needles can bend when entering tissue. Actual position must be verified by identifying
the echoes from the needle.
WARNING
Do not perform a needle procedure if the needle is not visible.
WARNING
Reverberation or other tissue artifacts may produce false needle images, which can cause
confusion in locating the actual needle image. Ensure that you are not using a false needle
image to locate the needle.