Changing the Configuration Set
3-54 Setting up your Monitor
Setting up your
Monitor
Remove the plug-in modules from the rack before starting the
procedure.
Ask your biomedical engineering department or Philips service
engineer to set Parameter Settings Transfer to OFF.
Procedure The current Configuration Set for the monitor is highlighted and the
universal settings are displayed at the top of the Task Window.
1. Press the softkey to select the Configuration
Set you require. The universal settings for the selected
Configuration Set are displayed.
2. Press the hardkey to switch to the selected set.
Wait at least one minute before plugging in or removing any modules
from the rack.
Change ConfSet
Confirm