oxyCRG Recordings
Recording Functions 6-37
Recording Functions
The oxyCRG is shown in three distinct non-overlapping “channels” on the
strip. The time, the three wave labels, and the scales for the oxygen and
beat-to-beat Heart Rate channels are printed periodically on the
recording.
Three speeds are available for oxyCRG recording: 1, 2, or 3 cm/min. You
can choose the speed in the oxyCRG Task Window
.
Note—oxyCRG recordings made at slow speeds (1 or 2 cm/min) can cause
deposits of paper debris on the print head over time. In this case,
recordings may appear unevenly faint. To correct, see “Cleaning the Print
head in the Plug-In Recorder” on page 6-52. Then reload the paper and
continue recording.
(8AMIN) 2 cm/min 05 JUL 9 11:20
05 JUL 94 11:20
HR 127
PULSE 125 ** PULSE 125 > 120
ABP 80/40 (59)
PAP -?- PAP ZERO+CHECK CAL
SpO2 87 **SpO2 87 < 90
NBP 82/41 (60) 05 JUL 94 10:56
AWRR 40
T1 35.2
T2 -?- T2
SpO2 80