Preset and configure the multimeter for AC voltage measurements with the
specified range and resolution. This command does not initiate the
measurement. For AC measurement, resolution is fixed at 6½ digits. Therefore
the resolution parameter only affects the front panel display.
CONFigure:CURRent:DC {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and configure the multimeter for DC current measurements with the
specified range and resolution. This command does not initiate the
measurement.
CONFigure:CURRent:AC {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and configure the multimeter for AC current measurements with the
specified range and resolution. This command does not initiate the
measurement. For AC measurement, resolution is fixed at 6½ digits. Therefore
the resolution parameter only affects the front panel display.
CONFigure:RESistance {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and configure the multimeter for 2-wire measurements with the
specified range and resolution. This command does not initiate the
measurement.
CONFigure:FRESistance {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and configure the multimeter for 4-wire measurements with the
specified range and resolution. This command does not initiate the
measurement.
CONFigure:FREQuency {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and configure the multimeter for frequency measurements with the
specified range and resolution. This command does not initiate the
measurement. For frequency measurements, the meter uses only one
“range” for all inputs between 3Hz and 300kHz. With no input signal applied,
frequency measurements return “0”.
CONFigure:PERiod {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and configure a period measurement with the specified range and
resolution. This command does not initiate the measurement. For period
measurements, the meter uses only one “range” for all inputs between 0.33