seconds and 3.3 µsec. With no input signal applied, period measurements return 0”.
CONFigure:CONTinuity
Preset and configure for a continuity measurement. This command does not
initiate the measurement. The range and resolution are fixed at 1kΩ and
digits respectively.
CONFigure:DIODe
Preset and configure for a diode measurement. This command does not initiate
the measurement. The range and resolution are fixed at 1Vdc with 1mA current
source and 5½ digits respectively.
CONFigure:TEMPerature
Preset and configure for a RTD measurement. This command does not initiate
the measurement. The resolution is fixed at 6½ digits.
CONFigure:TCOuple
Preset and configure for a thermocouple measurement. This command does not
initiate the measurement. The resolution is fixed at 6½ digits.
CONFigure?
Query a recent condition of function, and return a quoted string.
C.5 The Measurement Configuration Command
[SENSe:]FUNCtion “<function>”
Select a measurement function and enclose it in quotes in the command string
(FUNC “VOLT:DC”). Use one of the following strings.
VOLTage:DC VOLTage:AC VOLTage:DC:RATio
CURRent:DC CURRent:AC
RESistance (for 2-wire ohms) FRESistance (for 4-wire ohms)
FREQuency PERiod CONTinuity
DIODe TCOuple TEMPerature
[SENSe:]FUNCtion?
Query the measurement function and return a quoted string.
[SENSe:]<function>:RANGe {<range>|MINimum|MAXimum}