Query thermocouple sensor type.
[SENSe:]TCOuple:RJUNction:RSELect {REAL|SIMulated }
Select a reference junction type, real or simulated.
[SENSe:]TCOuple:RJUNction:RSELect?
Query the reference junction type, real or simulated.
[SENSe:]TCOuple:RJUNction:SIMulated {<value>|MINimum|MAXimum}
Set the default temperature of the simulated reference junction.
[SENSe:]TCOuple:RJUNction:SIMulated?
Query the default temperature of the simulated reference junction.
[SENSe:]TCOuple:RJUNction:REAL:OFFSet {<value>|MINimum|MAXimum}
Set the offset voltage of the real reference junction.
[SENSe:]TCOuple:RJUNction:REAL:OFFSet? [MINimum|MAXimum]
Query the offset voltage of the real reference junction.
[SENSe:]TEMPerature:RTD:TYPE {PT100|D100|F100|PT385|PT3916|USER|SPRTD|NTCT}
Set the RTD type.
[SENSe:]TEMPerature:RTD:TYPE?
Query the RTD type.
[SENSe:]TEMPerature:RTD:RZERo {<value>|MINimum|MAXimum}
Set the R-Zero constant for the RTD type.
[SENSe:]TEMPerature:RTD:RZERo? [MINimum|MAXimum]
Query the R-Zero constant for the set RTD type.
[SENSe:]TEMPerature:RTD:ALPHa {<value>|MINimum|MAXimum}
Set the alpha constant.
[SENSe:]TEMPerature:RTD:ALPHa? [MINimum|MAXimum]
Query the alpha constant for the set type.