Defaults
The default settings for Auto Zero are enabled. The user selected values
for Auto Zero are stored in a volatile memory and the default settings will
be restored when the meter is power-off.
How to set Auto Zero
You can change the Auto Zero setting through the front panel or through
the remote interface operation.
Front Panel Operation
The following steps show how to set Auto Zero directly through the front
panel.
Be aware that Auto Zero setting is always affected by the resolution
setting. Whenever the resolution is altered, Auto Zero may be changed
accordingly. The relation between resolution and Auto Zero is shown in
Table 4-1.
Press MENU and then ENTER on SET ADC. If “SET ADC” is not shown on
the display, use
and
buttons to locate it. Press ENTER on ZERO.
Again, use
and
buttons to locate it if it is not shown on the display.
Press ENTER on AUTO ZERO option and then use
and
to the “ON” or
“OFF”. Press “ENTER” by your choice. The locations of these buttons are
shown with red rectangle frames in Figure 4-1.
Procedures: MENU
SET ADC
ZERO
AUTO ZERO {ON|OFF}
Table 4-1
Integration Time(PLC)
Resolution
Auto
Zero
Remote Digits Displayed
DEFAULT
COMPATIBLE
DEFAULT
COMPATIBLE
0.02 0.02
Fast 4 ½ digits
Off 4 ½ 4 ½
0.1 1
Slow 4 ½ digits
On 4 ½ 5 ½
0.1 0.1
Fast 5 ½ digits
Off 5 ½ 5 ½
1 10
Slow 5 ½ digits
On 5 ½ 6 ½