64
6464
64
SENSe:VOLTage:DC:NPLCycles {0.02|0.1|1|10|MINimum|MAXimum}
SENSe:VOLTage:DC:NPLCycles? [MINimum|MAXimum]
SENSe:CURRent:DC:NPLCycles {0.02|0.1|1|10|MINimum|MAXimum}
SENSe:CURRent:DC:NPLCycles? [MINimum|MAXimum]
SENSe:RESistance:NPLCycles {0.02|0.1|1|10|MINimum|MAXimum}
SENSe:RESistance:NPLCycles? [MINimum|MAXimum]
SENSe:FRESistance:NPLCycles {0.02|0.1|1|10|MINimum|MAXimum}
SENSe:FRESistance:NPLCycles? [MINimum|MAXimum]
For frequency and period measurements, aperture time (or gate time) is
analogous to integration time, and you can use the following commands
to set it. Specify 10 ms (4.5 digits), 100 ms (default; 5.5 digits), or 1
second (6.5 digits).
SENSe:FREQuency:APERture {0.01|0.1|1|MINimum|MAXimum}
SENSe:FREQuency:APERture? [MINimum|MAXimum]
SENSe:PERiod:APERture {0.01|0.1|1|MINimum|MAXimum}
SENSe:PERiod:APERture? [MINimum|MAXimum]
The following table shows the relationship between the integration time
and the measurement resolution.
Integration
Time(PLC)
Resolution
DEFAULT COMPATIBLE
0.02 0.0001 x Full-Scale 0.0001 x Full-Scale
0.1 0.00001 x Full-Scale 0.00001 x Full-Scale
1 0.000001 x Full-Scale 0.000003 x Full-Scale
10 0.0000001 x Full-Scale 0.000001 x Full-Scale