13 CROSS-SECTIONAL EXPOSURE (OPTIONAL)
66 Planmeca ProOne User’s Manual
To adjust the height of the unit, press either of the height
adjusting buttons on the display until the chin rest is at the
level of the patient’s chin. Stretch and straighten the
patient’s neck.
The unit moves slowly at first, then faster.
Ask the patient to step forward, grasp the patient handles,
stretch up and bite the bite piece. The incisal edges of the
maxillary and mandibular teeth must be in the groove in
the bite piece.
NOTE
Patients are not allowed to hang on the patient handles.
The pull-down force applied to the patient handles may not
exceed 15 kg (33 lb).
NOTE
We recommend that you image patients with poor health
in a sitting position.
NOTE
If you are using the chin support, position the patient so
that the chin, just below the lower lip, touches the top bar.
NOTE
If you are using the chin support or chin cup, use for
example a cotton roll to ensure that the patient’s upper and
lower incisors do not overlap.
Close the temple supports by touching the temple support
button.