Setting Possible values
Starting
value
Increment Lower value Upper
value
Remark
[↳LIMIT]
1
0% 1% - 100% 100% Checktime limit.
No basic load,
only PID control
value
[BOOT
DELAY]
0 s 1 s 0 s 9999 s Control delay
period after the
start of the meas‐
uring point. After
it is switched on,
the unit only
measures but
does not control
during this period.
1 = in an upwards direction with mono-directional control: 0 ...+ 100% (setting with
PUMP: dosing↑), in a downwards direction: - 100 ... 0% (setting with PUMP: dosing↓).
2 = When switching over the metering direction, all actuators in the controller are reset to
the factory settings for the selected metering direction.
Operating menus
68