Setting Possible values
Display Starting
value
Increment Lower
value
Upper
value
Remark
[↳TREF]
25 °C 1 15 °C 30 °C Reference tempera‐
ture
[TCOEFF]
1.9 %/°C 0.1 0 %/°C 9.99 %/
°C
Temperature coeffi‐
cient
[FILTER]
3 s 1 3 s 30 s Measured value fil‐
tering in seconds
1
[TDS]
0.64 0.001 0.004 1.000 TDS conversion factor
[TEMP]
auto man Correction value
source (Pt100 (0),
manual)
auto
[↳ UNIT]
°C °C
°F
Correction value unit
[↳ VALU]
25.0 °C 0.1 0.0 °C 150.0 °C Manual correction
value °C
[↳
VALUE]
77.0 °F 0.1 32.0 °F 302.0 °F Manual correction
value °F
[CON‐
TACT]
pause pause Configuration of digital
contact input
hold
[↳ POL]
norm.open, norm.open, Polarity of the contact
input
norm.closed
[↳ DELAY
OFF]
0 s 1 0 s 3600 s Switch-off delay of
contact input
[↳
ALARM]
off on Alarm at
[Hold]
or
[ PAUSE]
event
off
1)
[FILTER]
: The default value of 3 seconds is suitable in most cases. You should only
increase the default value of 3 seconds with fluctuating readings, a factor that also
increases the setting time of the display value.
Operating menus
71