Configuration
163
8.2.3 Default settings
The “Default settings” screen is opened by highlighting the ‘Default settings’ icon use
the navigation keys then pressing
e to confirm selection. Alternatively, pressing 3
while in the configuration screen to directly access the available settings. The screen is
used to set specific default parameters. The default changes are effective upon restart
or if a new file is created.
‘Default RPM’ – used to set the required default rpm. Use
f/g and
highlight ‘Default RPM’ then use the data entry keys to enter the required
default rpm. Confirm entry by pressing either
e or ß. Every new alignment
file created will then automatically be preconfigured with this RPM value already
entered.
‘Reference diameter’ – used to set the required reference diameter. Use
f/g and highlight ‘Ref. diameter’ then use the data entry keys to enter
the required reference diameter. Confirm entry by pressing either
e or ß.
Every new alignment file created will then automatically be configured with this
coupling diameter value already entered.
‘Default sensor’ – The system’s default sensor is the RS5 BT sensor. Depending
on the system configuration, the default sensor may be set to RS5 sensor or
ROTALIGN sensor or OPTALIGN sensor.