Page 37
8) Run FlowPulse PC (do not click connect), go to System menu and
select Bootloader Control
9) On the Flow Pulse Bootloader window, select the Com Port number
and click Connect
10) Click Load Hex File and choose the relevant firmware file for the
Handheld.
11) Click Erase-Program-Verify”, and wait for completion when a
Verification Successful message will be displayed, then click Run
Firmware
12) Disconnect serial lead and then reassemble the unit and screws ensuring
that all rubber washers or seals are in place
13) Power-up the unit and press F4Device Info to verify the new
firmware version number is correct.