Loading...
EM510 SENSe Subsystem:GCONtrol
4065.7763.32-01.00 5.84
. . . . [:INCRement]? [MINimum|MAXimum]
Query about the MGC stepwidth.
Parameters:
none query about the currently set stepwidth
MINimum|MAXimum query about the smallest|largest stepwidth
Result:
MGC stepwidth in dB
Example:
GCONtrol:STEP? -> 10
. . :MODE FIXed|MGC|AUTO|AGC
Type of gain control.
Parameters:
FIXed|MGC control is determined by MGC value
AUTO|AGC control is automatically generated (AGC)
*RST state:
AUTO
Example:
GCONtrol:MODE AUTO
. . :MODE?
Query about the type of gain control.
Parameters:
none
Result:
FIX, AUTO see above
Example:
GCONtrol:MODE? -> AUTO