SENSe Subsystem :MSCan EM510
5.85 4065.7763.32-01.00
. :MSCan
The MSCan system controls the frequency function of the device, provided the memory scan has been
activated by SENSe:FREQuency:MODE MSCan. Each scan is started by INITiate[:IMMediate].
The memory locations are placed in the MEMory subsystem and are set for query during the scan.
. . :CHANnel <name>
Setting the current memory location.
The memory locations are addressed by the text (see "Parameters" on page 5.9) MEM0 to MEM9999
(memory location 0 to memory location 9999) and the next free memory location by ACTUAL .
During the memory scan, this command is not permitted.
Parameters:
<name> MEM0 to MEM9999 | NEXT
*RST state:
MEM0
Example:
MSCan:CHANnel MEM357
. . :CHANnel?
Output of current memory location.
Parameters:
none
Result:
Number of current memory location.
Example:
MSCan:CHANnel? 357
. . :CONTrol
. . . :[ON]<control_function> {,<control_function>}
Command for switch-on of the 'STOP:SIGNal'functions.
With 'STOP:SIGNal' the disappearance of the signal during the dwell time causes the dwell time to
be aborted. The hold time after the disappearance of the signal is set with
SENSe:MSCan:HOLD:TIME.
Parameters:
<control_function> is the following string:
'STOP:SIGNal' switches the signal-controlled dwell time on
*RST state
After *RST there is no control function running
Example:
MSCan:CONTrol 'STOP:SIGN'