SENSe Subsystem :SWEep EM510
5.93 4065.7763.32-01.00
. :SWEep
The SWEep system controls the frequency function of the device if the frequency scan has been
activated by the SENSe:FREQuency:MODE SWEep command. Each scan is initiated by
INITiate[:IMMediate].
. . :CONTrol
. . . :[ON] <control _function> {,<control_function>}
Command for switch-on of the STOP:SIGNalfunctions.
With "STOP:SIGNal" the disappearance of the signal during the dwell time causes the dwell time to be
aborted. The hold time after the disappearance of the signal is set with SENSe:SWEep:HOLD:TIME.
Parameters:
<control_function> is one of the following strings:
"STOP:SIGNal" switch-on signal-controlled scan continuation
*RST state:
After *RST no control function is switched on.
Example:
SWEep:CONTrol "STOP:SIGN"
. . . [:ON]?
Query about the switched-on scan-control functions.
Parameters:
none
Result:
List of switched-off control functions. If no function is active, a zero string ("") is output.
The following strings are to be expected:
"STOP:SIGN" signal-controlled scan continuation switched on
"" zero string : none of the two functions is active
Example:
SWEep:CONTrol? -> "STOP:SIGN"