EM510 SENSe Subsystem :SWEep
4065.7763.32-01.00 5.94
. . . :OFF <control_function>{,<control_function}
Switch-off of one or several scan-control functions.
Parameters:
see SENSe:SWEEp:CONTrol [:ON]
*RST state:
after *RST there is no control function running
Example:
SWEep:CONTrol:OFF "STOP:SIGN"
. . . :OFF ?
Query about the switched-off scan-control functions.
Parameters:
none
Result:
List of switched-off control functions. Strings see SENSe:SWEep:CONTrol [:ON]
Example:
SWEep:CONTrol:OFF? ->"STOP:SIGN"
. . :COUNt <numeric_value>|MINimum|MAXimum|INFinity
Information about the number of sweeps.
Parameters:
<numeric_value> number of sweeps
MINimum|MAXimum minimum|maximum number
INFinity infinite number
*RST state:
INFinity
Example:
SWEep:COUNt 100