EasyManua.ls Logo

R&S NRP8S - 6 Browser-Based User Interface (Continued)

R&S NRP8S
184 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Browser-Based User Interface
R&S
®
NRPxxS(N)
46User Manual 1177.5079.02 ─ 10
End Exclude
Sets a time that is to be excluded at the end of the measurement period.
Remote command:
[SENSe<Sensor>:]TIMing:EXCLude:STOP on page 109
Trigger Level
See "Trigger Level" on page 54.
Dropout Tolerance
Sets the dropout time. The dropout time is a time interval in which the pulse end is only
recognized if the signal level no longer exceeds the trigger level.
Remote command:
[SENSe<Sensor>:][POWer:]BURSt:DTOLerance on page 97
6.4.3 Timeslot Mode
Describes the parameters of the timeslot average measurement.
Further information:
Chapter 9.7.3, "Timeslot Average Measurement", on page 97
Detailed description of the timeslot average mode and its remote commands
Access: main dialog of the web user interface > navigation pane > "Timeslot"
Number of Timeslots..................................................................................................... 46
Nominal Width...............................................................................................................47
Start Exclude.................................................................................................................47
End Exclude..................................................................................................................47
Trigger Source...............................................................................................................47
Trigger Level................................................................................................................. 47
Number of Timeslots
Sets the number of simultaneously measured timeslots. Up to eight slots can be
selected.
Measurement Modes

Table of Contents

Other manuals for R&S NRP8S

Related product manuals