EasyManua.ls Logo

RAUMEDIC MPR2 logO - Page 61

Default Icon
99 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instructions for Use RAUMEDIC
MPR2 logO
Manufacturer: zwo-400EN
RAUMEDIC AG, Hermann-Staudinger-Strasse 2, 95233 Helmbrechts, Germany Rev. 9 2020-09-22
Page 61 of 99
Step 1:
Call device setup
and select
special function OutpO
Step 2:
Make the cable connections as
described and then press the OK
key
Step 3:
Carry out a null measurement at
the pressure inputs of the external
system. Press the OK key once
the null measurement has
successfully completed for both
pressure inputs
Step 4:
A control signal of 20 mmHg is
output on both
analogue outputs.
Check the displays on the external
system and confirm
by pressing
OK.
Step 5:
After the OK confirmation in step
4, enter device setup (figure 4).
Exit screen 4 by pressing the next
key
Step 6:
pO2 is forwarded via Out1 and
ICP2 via Out2.
This document is for information only

Table of Contents

Related product manuals