Example: Failure of the temperature channel a fixed temperature compensation is carried out at
37 °Celsius, patient temperature 39 °Celsius, displayed value pO2 = 20.0 mmHg. The error at this point is
-1 mmHg, i.e. the measured value is 19.0 mmHg.
If the temperature measurement T2 has failed or is defective, the measurement of pO2 continues
to be possible in a temperature range of 32 °Celsius to 42 °Celsius.
In the measurement of very small physiological partial oxygen pressures an additional error of -
0.5 mmHg to +0.3 mmHg occurs (relative to the specific accuracy, cf. 11.1 Technical data
).
4.4.4 Transducer for temperature measurement
Please observe chapter 3.3 Measurement of the body temperature.
Use approved accessories only.
The MPR2 has two temperature input ports T1 and P2/T2 to be used to connect temperature sensors (port
T1: temperature sensors made by Exacon or RAUMEDIC multi-parameter catheters with integrated
temperature measurements; port P2/T2: multi-purpose port to be used for the connection of a PTO
catheter). A ¼ inch jack plug can be used to connect temperature sensors to port T1. The PTO cable is
required to connect a PTO catheter to the port P2/T2.
The temperature sensor connected is recognised automatically. The same applies corresponding, if you
want to connect another temperature sensor. If the temperature to be measured is within a range of
15 °Celsius < T < 45 °Celsius, the temperature will be displayed in the measured value field of the overview
screen .
Please keep in mind that apart from the catheter, the ICP-Temp cable, (article No. 094328-001),
and the ICP-Temp separator with jack plug, (article No. 094323-001), are required when the
RAUMEDIC multi-parameter catheters with temperature measurement are used.
Please observe that when RAUMEDIC PTO catheters are used for temperature measurement, the
PTO cable, (article No. 095624-001), is required in addition to the NEUROVENT-PTO catheter,
(article No. 095008-001).