3 Patient monitoring
Note: This chapter describes the units and their functions as well as the extension options of the MPR2.
A detailed description of the operation of the MPR2 is rendered in chapter 4 Operation of the
MPR2, and the technical data are listed in chapter 11.1 Technical data.
3.1 Invasive pressure measurement IBP
3.1.1 General aspects
Applied part type CF
Type: for pressure sensors with radiometric characteristic 5 µV / V / mmHg
Number: 2
Where: Figure 2: Connection field, connection ports [20] and [21]
Application: see below
The MPR2 can be used to measure and display up to two invasive pressures. The pressure channels are
type CF and have been classified as protected against defibrillation.
Invasive procedures comprise risks for the patients. Apply aseptic techniques, and observe the
instructions of the catheter manufacturer.
Use accessories only which are listed in the overview in chapter 9 Accessories
. No splitting into spatially
separate sensors for pressure and temperature may be effected at the multi-purpose port P2/T2.
Use transducers and cables only which have been protected against defibrillation. Observe the
instructions of the transducer and/or catheter manufacturer.
Disposable transducers are intended for one-time use only.
The MPR2 has two input sockets P1, P2/T2, to which up to two transducers or micro-chip precision pressure
catheters can be connected. By means of the software, pressure measurement inputs can be allocated to
different measuring locations.
Designation Definition
ART arterial pressure
CVP central venal pressure
ICP1 intra-cranial pressure 1
ICP2 intra-cranial pressure 2
GP1 standard label - pressure 1
GP2 standard label - pressure 2
GP3 standard label - pressure 3
Table 3: Designation of the measuring location
The allocation of the measuring location of the pressure channel sets its scale, alarm source and alarm
limits.
Specification of the pressure channels: cf. chapter 11 Technical parameters (Specification)
Recommended accessories: cf. chapter 9 Accessories