Responding to Alarms
42
If this light is
on
And this condition
exists
It means...
Loose Lead Continuous yellow
light and continuous
alarm
The yellow loose lead light and the alarm may sound
continuously when there is a problem with any of the following:
Lead wires
Electrodes
Electrode belt
Patient cable, or
Connections between the baby’s skin and the electrodes,
the lead wires, the patient cable, and the SmartMonitor 2.
Loose Lead Continuous yellow
light and no alarm
If you correct the problem, the alarm will stop. However, the
yellow light remains on until you press the RESET button.
Power Continuous green
light, continuous
alarm, with no other
lights lit.
Check the display for error messages. If no error messages
are displayed, the SmartMonitor 2 was turned off improperly
causing a sibling alarm.
To resolve:
Press and hold the blue RESET button.
Press and release the gray POWER button.
Wait 2 seconds then release the RESET button.
All All lights are blinking
and the alarm will
come on for 3 seconds
and then off for 1
second.
Check the display for error messages. If there is an error
message, enter it on your log sheet.
Turn the monitor off and then back on. If the monitor functions
normally continue to use the monitor.