Portable Operation of the SmartMonitor 2
50
SmartMonitor 2 Battery Pack
Warnings
A warning indicates a potentially harmful situation.
Never change the battery pack while the battery charger is plugged in and/or the SmartMonitor 2 is
being operated.
Do not operate the SmartMonitor 2 with the battery pack removed (and the Battery Charger
connected). There is no power loss alarm feature under these conditions.
Do not disassemble the battery pack or batteries. The batteries contain electrolytes, which can cause
injury to eyes, skin, and clothing.
Do not dispose of the old battery in fire or incinerator. The battery pack contains ingredients that
could be explosive and cause serious personal injury.
Cautions
A caution indicates a condition that may lead to equipment damage, malfunction, or inaccurate operation.
The battery pack has a locking connector. Do not attempt to remove the battery pack without
releasing the latch on the connector. Doing so will damage the battery wires.
Use only Respironics-supplied battery packs. Use of other battery packs could cause malfunction.