Responding to Alarms
41
Responding to System Alarms
A System Alarm indicates that the monitor may not be functioning properly or at optimum capacity. The
information in this section will help you respond appropriately to system alarms. When a SmartMonitor 2
system alarm occurs, one of the lights at the bottom of the front panel will come on.
If this light is
on
And this condition
exists
It means...
Power
Continuous green
light, no alarm.
Normal Operation. The green power indicator light will come
on and stay on for as long as the monitor is on.
Charger
Continuous green
light, no alarm.
Normal Operation. The green charger light will come on and
stay on while the battery charger is plugged into an active
outlet and connected to the SmartMonitor 2.
Low Battery
Flashing yellow light,
continuous alarm.
Continuous yellow
light, continuous
alarm.
This is a warning that the battery voltage is very low and
should be recharged soon. (See Charging the Battery in this
manual). Press the RESET button to temporarily silence the
alarm. The alarm will resound in 2 minutes if the monitor has
not been plugged in. The yellow light will continue to flash.
This is a warning that the battery is too low for the
SmartMonitor 2 to operate properly. The SmartMonitor 2 must
be recharged. Turn the monitor off. Then, recharge the
battery. (See Charging the Battery in this manual).
If you do not recharge the battery, the system will
automatically shut down.
Memory Full Flashing yellow light,
continuous alarm
When the SmartMonitor 2’s Memory Almost full parameter is
reached, the Memory Full light will flash. The alarm will sound
continuously. The alarm will sound only if the dealer programs
the SmartMonitor 2 to do so at the 50% full or at 80% full.
Press the RESET button to silence the alarm. The light will
blink every second.
Memory full is a warning condition. You can continue
monitoring. The caregiver should contact the dealer to
resolve the alarm.
Memory Full Continuous yellow
light, continuous alarm
SmartMonitor 2 memory is 100% full. Press the RESET button
to silence the alarm.
The alarm will sound only if the dealer programs the
SmartMonitor 2 to do so.
The light will stay on continuously. The caregiver should
contact the dealer to resolve the alarm. (See Transferring the
Monitor Memory in this manual for more information).