12
GA--A 202
3Setup
WARNING!
The device is not protected against explosions.
Explosion hazard.
Do not operate this device in areas where there is the danger of
explosion.
WARNING!
Danger if a power supply without protective earth is used.
Danger of electric shock!
Connect the device only to a power supply with protective earth.
.
NOTE!
Check that the mains/line voltage is the same as the voltage specified on
the identification plate. Connect the device only with the supplied power
cable o r a power cable meeting the same specifications.
Make sure that the vent slots are not blocked.
CAUTION!
Danger of faults and malfunctions.
To guarantee the safety of the u ser, the patient and others use only
accessories and spare parts specified by the manufacturer of this
product.
Other accessories or spare parts can cause the emission of
increased electromagnetic radiation or reduced immunity against
interference.
.
IMPORTANT!
Medical devices are subject to special precautions with regard to electro-
magnetic compatibility (EMC).
Make sure you observe the notes on EMC for installation and operation.
Medical electrical devices can be influenced by mobile HF communication
devices.
If it is necessary to stack the devices or place them next to each other
and HF interference is observed, make sure you observe the intended
use of the devices.
3.1 Preparation of POWER CONTROL
Z Connect device to mains/power.
Z Connect footswitch cable to device.
Z Switch on the power switch.
.
NOTE:
To disconnect the cable from the device, pull back locking sleeve (B) on
connector (front portion with red marking).
B