GA--A 202
7 Reprocessing and maintenance
7.1 Reprocessing of device
WARNING!
Make sure that no humidity enters the device.
Danger of electric shock.
Before reprocessing switch off the device and disconnect it from the
power supply.
Clean the device with a soft cloth moistened with surface disinfectant,
alcohol or spirit.
Follow the disinfectant manufacturer's instructions.
.
NOTE!
For a description of the surface cleaning and disinfecting agents for devi
ces and universal device carts / trolleys and utility carts approved by Ri
chard Wolf, please refer to manual GA-J020 "Reprocessing of RICHARD
WOLF Heat-stable Instruments".
.
IMPORTANT!
Make sure that no humidity enters the device. Do not use any cleaning
agents, scouring agents or solvents on this device.
7.2 Maintenance Device
.
IMPORTANT!
In your inquiries or correspondence please always indicate the model and
serial number on the identification plate. Further documentation is avail
able from the manufacturer on request.
7.2.1 Maintenance intervals
.
IMPORTANT!
To prevent incidents caused by aging and wear, the device and its ac
cessories must be serviced at adequate intervals. Depending on the fre
quency of use, however at least once a year, an expert must check the
device and carry out a safety test.