5.3 Operation of device
CAUTION!
Caution in the case of poor visibility.
Tissue in the operating area may inadvertently be caught and
damaged.
Operate the blade only if the cutting window is easily visible.
CAUTION!
Do not resharpen the rotary blades / abraders!
Operate the rotary blades / abraders only in liquid!
Ensure sufficient suction to cool the rotary blades / abraders and to re
move the ablated tissue from the cutting area.
CAUTION!
Rotating cutting edges of the inner blades / abraders must not touch
metal parts (e.g. trocar sleeves) or arthroscopes,
as this may damage both instruments.
In the case of such contact, carry out a check for damage and complete
ness immediately. Extract missing parts.
Do not use the instruments if damage such as deformed cutting edges,
cracks or fractures are visible.
Make sure that missing instrument parts do not remain in the
patient.
.
IMPORTANT!
Follow the instructions on interval operation, as otherwise the grip surface
of POWER STICK M4 can become very hot.
.
IMPORTANT!
The inner and outer blades of a set must not be exchanged with the
blades of other sets.
.
NOTE!
An increased suction rate has a positive impact on cutting performance
and reduces the danger of clogging.