7.3 Reprocessing of POWER STICK M4 (8564.121)
.
IMPORTANT!
Further notes and instructions on reprocessing are described in manual GA-J 020
"Reprocessing of RICHARD WOLF Heat-Stable Instruments", and these must be
followed.
7.3.1 Manual reprocessing
Wet preparation at the point of use:
Z Immediately after use, immerse all parts seperately in a combined dis
infecting and cleaning solution (disconnect connection cable and re
move valve).
' This prevents microorganisms from being transmitted and contamin
ant residues from drying on the instrument.
Disassembly before cleaning:
Z Push out valve (14.2) laterally, while moving the valve back and forth
between “ON" and “OFF".
Z Disconnect the connection cable.
Manual cleaning:
Z Carefully rinse all parts (POWER STICK, connection cable and valve)
first with cold and then warm water under the tap.
Z Immerse all parts in a combined cleaning and disinfecting solution.
Clean hollow spaces (e.g. channels) with suitable brushes (while the
parts are immersed).
Z Rinseoutthesuctionchannelwithopensuctionvalveusingacleaning
gun(seefigure).
Z Clean the inside blade retaining bolt with a cleaning gun (see figure).
Z Thoroughly rinse with demineralized water, then dry inside and out with
acloth,swaborcompressedair.
14.2